Electronic personal dosimeters of PM1610 series provide measurement of dose equivalent (DE) and dose equivalent rate (DER) of continuous and pulsed X-ray and gamma radiation in the energy range from 20 keV to 10 MeV, as well as time of dose accumulation. Instruments have audible, visual and vibration alarms to warn the user when preset dose/dose rate threshold levels are exceeded. Shockproof rubberized case and LCD with fluorescent backlight allows operation in harsh environment. Dosimeters are compact and easy in use, have long life built-in rechargeable battery.
PM1610 series is recommended for use by medical professionals, first responders, customs and border patrol, radiological and radioisotope laboratories, professionals exposed to pulse radiation.
Dosimeters allow setting of two DER alarm threshold levels and two DE alarm threshold levels. Instruments continuously control the threshold levels and alert the user with audible, light and vibration alarms when the threshold levels are exceeded. Instruments automatically record data into permanent memory and store up to 7500 events of DER and DE measurement history in non-volatile memory. PM1610-01 modification is equipped with data communication channel (ISO15693 standard).
PM1610 - basic model
PM1610-01 - modification with data communication channel (ISO15693 standard)
PM1610A - modification with extended DE measurement range, relative DER measurement error
PM1610A-01 - modification with extended DE measurement range, relative DER measurement error and data communication channel (ISO15693 standard)
Personal Dose Tracker System developed by Polimaster, enables connection up to 100 units of PM1610. PDT System is intended for continuous account for external radiation exposure, maintenance of dose history database and calculation of the personnel accumulated doses (individual as well as group doses) during the set period of time, alarm when preset thresholds (individual as well as group thresholds) are exceeded. Software enables information transfer by USB-protocol to a personal computer to analyze, process, and generate corresponding databases in a control center or an expert center.
Instrument battery will gradually run down during a long-term storage (even if the instrument is switched off). Over-discharged batteries could lose some capacity. For this reason Polimaster recommends users to charge the battery periodically (at least once per six months).
Designed to meet ANSI N42.20 - 2006 and IEC 61526: 2010 requirements
Complies with ANSI 42.34, section 9.11 requirements by the level of emitted radio-noise
Detector | Geiger-Muller tube | ||||||||
Dose equivalent rate (DER) measurement range | 0.1 µSv/h to 10.0 Sv/h | ||||||||
DER indication range | 0.01 µSv/h to 12.0 Sv/h | ||||||||
DER measurement accuracy | ±(15+0.0015/Ḣ)% (in range 0.1 µSv/h - 10 Sv/h, where Ḣ - dose rate in mSv/h) | ||||||||
DE measurement range:
|
|||||||||
DE indication range | 0.001 µSv to 12.0 Sv | ||||||||
DE measurement accuracy | ± 20 % | ||||||||
Energy range | 0.02 to 10.0 MeV | ||||||||
Energy response relative to 0.662 MeV:
|
|||||||||
PC communication | USB | ||||||||
Data exchange | ISO15693 interface | ||||||||
Power supply | built-in rechargeable battery (502030 240mAh MOBILE STAR) | ||||||||
Battery lifetime:
|
|||||||||
Ingress protection | IP65 | ||||||||
Drop test on concrete floor | 1.5 m onto hard wooden surface | ||||||||
Dimensions | 58 x 59 x 20 mm (2 2/7'' x 2 2/7'' x 5/7'') | ||||||||
Weight | no more 0.07 kg (2.5 oz) | ||||||||
Alarm Type | visual, audio, vibration | ||||||||
Mean service lifetime | 10 years | ||||||||
Mean recovery time | 60 min | ||||||||
Operating temperature | -20 °С to + 50 °С | ||||||||
Humidity | 98 % at 35 °С | ||||||||
Atmospheric pressure | 84 to 106.7 kPa |